成人亚洲精品777777-成人亚洲精品久久久久-成人亚洲精品久久久久软件-成人亚洲区无码区在线点播-成人亚洲性情网站www在线观看-成人艳情一二三区

Products

current position:home page>Products>Analysis Instrument>PST>

Binary Gas Analyzer for Hydrogen Monitoring - Michell XTC601

Updated:2025-03-13

Views:200

  • Brand: PST/Michell
  • Model: XTC601
  • Description: A robust, linear and stable thermal conductivity analyzer for measurement of binary gas mixes such as air in hydrogen, nitrogen, argon, helium or carbon-dioxide. The sensor is housed in a rugged casin
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

Binary Gas Analyzer for Hydrogen Monitoring - Michell XTC601 Description:

User's Manual Datasheet

Thermal Conductivity Technology in Safe or Hazardous Areas

Features

Suitable for use in ATEX, UKCA, IECEx, TR CU Ex & cQPSus certified Hazardous Areas

Optional status LEDs following NAMUR standard

Compact and rugged design with an EExd enclosure and 4...20 mA output

Touch-screen display allows calibration or adjustment without the need for a hot works permit

Accuracy of better than ±2 % full scale

IP66 rated enclosure

Low cost of ownership due to minimal maintenance

Typical ranges available: 0...5 % up to 0...100 % (others gas dependent)

Stability ±0.5 % per month (of range)

A robust, linear and stable thermal conductivity analyzer for measurement of binary gas mixes such as air in hydrogen, nitrogen, argon, helium or carbon-dioxide. The sensor is housed in a rugged casing, making it suitable for a wide range of applications. The analyzer can be supplied explosion-proof and suitable for hazardous areas.

Specifications

Performance
Measurement Technology Thermal Conductivity sensor
Measured Gases Air, Ar, CH4, CO2, H2, He or N2
Gas Requirements Non-condensing sample with particles <3μm
Measurement Range Selectable from 0...5 up to 0...100 %
50...100 % up to 90...100 %
Measurement Range (H2 & He) Selectable from 0...1 % up to 0...100 %
50...100 % up to 99...100 %*
Display Resolution 0.1 %
Display Type Backlit LCD
Intrinsic Error (Accuracy) < ±2 % of range**
Response Time (T90) Standard < 20 seconds**
Repeatability ±0.2 % of range
Linearity ±1 % range
Zero Stability ±0.5 % of range per month
Range Stability ±0.5 % of range per month
Sample Flow Rate 100...500 ml/min (0.2...1.27 scfh)
Sample Flow Rate(Ex) 270...330 ml/min (0.54...0.66 scfh))
Sample Flow Effect
(calibrated at 300 ml/min)
< 1 % of range for flows:
100...500 ml/min
Sample Pressure Analyzers are callibrated at atmospheric pressure and must be operated with an atmospheric vent, unless otherwise stated
Maximum Safe Pressure 4 barg (58 psig)
Sample Temperature A constant temperature of +5...+45/+55 °C (+41...+113/+131 °F)
Sample Cell Temperature +50/+60 °C (+122/+140 °F)
Electrical Specifications
Analog Inputs 2 off 4...20 mA inputs
One for an external sensor that can be displayed on the screen
One to act as an active compensation for the process conditions
Analog Outputs 2 off 4...20 mA outputs
Output Ranges Primary range is set to the calibrated range of the instrument
The second is user selectable
Alarms 2 off single pole changeover (SPCO) relays (250 V, 5 A max)
Datalogging The user can use the digital communications to log the output from the analyzer
The unit will store 40 alarm points and the min/max O2 concentrations with date and time stamp
Digital Communications Modbus RTU over RS485
Power Supply 24 V DC; 1.0 A max
Operating Conditions
Ambient Temperature +5...+60 °C (+41 °F...+140 °F)
+5...+50 °C (+41...+122 °F) cQPSus
Mechanical Specifications
Warm Up Time < 25 minutes
Stabilization Time 5 minutes
Dimensions 234 x 234 x 172 mm (w x d x h)
Weight 9.7 kg
Wetted Materials 316 stainless steel, boroscillicate glass, platinum, (plus O-ring)
O-Ring Materials Viton, Silicone or Ekraz
Gas Connection 1/4" NPT female (GP 1), 1/8" NPT female (Ex & GP2)
Ingress Protection IP66, NEMA 4
Hazardous Area Classification
ATEX/UKCA II 2 G D
Ex db IIB +H2 T6 Gb
Ex tb IIIC T85 °C Db IP66
IECEx Ex db IIB +H2 T6 Gb
Ex tb IIIC T85°C Db IP66
Temperature ranges for ATEX/UKCA and IECEx as per O-ring type Silicon: Ta = -40 °C...+60 °C (+50 °C for N. America, +55 °C for TR CU Ex (EAC))
Viton: Ta = -15 °C...+60 °C (+50 °C for N. America, +55 °C for TR CU Ex (EAC))
Ekraz: Ta = -10 °C...+60 °C (+50 °C for N. America, +55 °C for TR CU Ex (EAC))
cQPSus Class I, Division 1, Groups B,C,D T6
Class I, Zone 1
AEx db IIB+H2 T6 Gb / Ex db IIB+H2 T6 Gb
TR CU Ex 1Ex d IIB+H2 T6 Gb
*Ranges 98-100 % and 99-100 % (for H2 and HE only) will require a flowing reference and selection of the 4-port model.
reference and selection of the 4-port model.
**The typical intrinsic error and response time will be as above. Certain gas combinations and/or ranges may have different specifications. Please consult Michell Instruments for specific cases. For example, CO2 / N2 would be < 50 seconds for T90.

Note: If the 4-port version is selected then a reference gas with a flow rate of 300 ml/min is required

Suitable Applications

Hydrogen coolant in electricity turbines

Product quality in air separation plants

Syngas production

Helium recovery

Fuel cell research

Product quality such as Air in Argon for double glazing

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@dorgean.com
主站蜘蛛池模板: 8008.榴莲 | 色婷婷中文网 | 亚洲视频在线观看不卡 | 欧美日韩一区二区在线 | 国产激情偷乱视频一区二区三区 | www.亚洲一区| 国内精品久久久久久影院网站小说 | 亚洲视频久久 | 国产在线综合视频 | 日韩1区 | 欧美精品一区在线看 | 小说区图片区激情区视频区 | 国产高清视频一区二区 | 久久两性视频 | 26uuu色噜噜欧美在线播放 | 亚洲免费视频一区二区 | 99热这里只有精品免费 | 亚洲精品入口一区二区在线播放 | 久操视频在线免费观看 | 竹内纱里奈55在线观看 | 99久久久久久久久 | 亚洲第一视频在线观看 | 日韩成人在线影院 | 久久青青| 精品一区二区三区四区乱码 | 天天操夜夜干 | 亚洲99精品A片久久久久久 | 精品视频免费 | 久久精品综合国产二区 | 欧美三级不卡 | 黄色片免费视频 | 久久久久久片 | 亚洲网站在线 | 国产一区二区三区四区hd | 国产福利一区视频 | 九九热免费在线视频 | 激情婷婷在线 | 久久综合九色综合亚洲小说 | 欧美福利视频 | 久草成人| 麻豆一区 |